Conforms to |
BS EN ISO 7396-1:2016, HTM 02-01 Part A & BS EN 60601–1 |
Class IIb – MDR 2017 / 745 |
Body |
Fire retardant dust-proof CRCA 18 SWG steel enclosure |
Finish |
Powder coating – Enclosure RAL 9003 |
Mounting type |
Horizontal – Surface mountable |
Enclosure dimensions |
2&3 Gas module – 210x204x80 and 4,5 & 6 Gas – 303x204x80 (L x H x T in mm) |
Gas services |
2, 3, 4, 5 & 6 Gas including Vacuum |
Gas & Vacuum connections |
½ inch BSP 20mm long on the top (Riser) for connecting network gas services. |
Power connection |
External 240V AC (50 Hz) to 12V DC adapter for audio visual display |
Warning system |
Microprocessor based warning module should be adjustable in-situ to set operational parameters and digital pressure unit in psi / bar or kg/cm2 |
Display & Audio signal |
Digital & illuminated LED display for “Normal”, “High”& “Low” with High/low alarm set- points for each gas (Low only for Vacuum) service |
Flashing LED display coupled with audio warning in case of preset deviation with the provision to “Mute” audio signal for 15 minutes max. |
Minimum size of 20 x 14 mm bright LED signals for clear visibility from distance. |
ABP Sensors |
Accurate, calibrated & temperature compensated PCB mount Pressure Sensors |
a) Sensing pressure span: 60m bar to 10 bar | 1 psi to 150 psi Digital Output |
b) Temperature span: 0°C to 50°C (32°F to 122°F) for gases (-)20°C to 85°C {(-)4°F to 185°F} for Vacuum. |
c) Accuracy: 0.25% of FSS BFSL |
Information & Control |
The information of alarm, current status & test conditions through “Touch button” |
Display & Audio blocker |
Provision to block display & audio signal of a particular supply line for long service period is built in. |
Earthing |
Protective earth |
Protection |
Class II & IP 54 |
Certifications |
CE Medical Device Marking according to 93/42/EEC Directive |
ISO 13485:2016 – Medical devices – Quality Management System |