RESPIRATORY THERAPY- BPC FLOWMETER WITH HUMIDIFIER (0-15 LPM)

Conforms to: IS / ISO : 15002: 2008 + Amd.1: 2019:,
IS / ISO : 7396 – 1 : 2016, HTM 02 – 01
Class IIb – MDR 2017 / 745

Certifications: ISO 9001:2015 – Quality Management Systems
ISO 13485:2016 – Medical Devices – Quality Management Systems
CE Medical Devices Marking according to 93/42/EEC Directive

Essentia medical gas flowmeters are designed to measure and regulate flow. These are calibrated to specific gas & flow range. Oxygen flowmeter is intended for use with various delivery system when bubble humidifier is attached downstream.

Humidifier prevents drying of airway passage associated with the inhalation of oxygen which is inserted in the inspiratory circuit.

These flowmeters are widely used in home care, health centers, district / general / regional hospitals, tertiary care, emergency vehicle etc.

Three models are available:
1. ESSENTIA OXY ADULT – 0 to 15 lpm (For Adult use)
2. ESSENTIA OXY BABY – 0 to 3 lpm (For Pediatric use)
3. ESSENTIA AIR – 0 to 15 lpm (For Adult use)

Features

  • Back pressure compensated design to give specified accuracy for entire range.
  • Standard Thorpe style for medical Oxygen (Adult & Paediatric) & Medical Air
  • Step-less control
  • Chromium plated brass body with SS 316 wire mesh inlet filter
  • Durable impact resistant Polycarbonate measuring and outer tube.
  • Transparent, clearly readable and permanent graduated column.
  • Adaptable to Quick or DISS probes / connectors
  • Standard 3/8” BSP RH with option of 1/8” BSPT connection.
  • Poly-carbonate humidifier with auto-reset table pressure relief valve for safety.
  • Suitable for MGPS or through regulator with steady 3.5 bar outlet pressure

Technical Information:

MODELSEssentia Oxy AdultEssentia Oxy BabyEssentia Air
Flow (L / Min) 0 – 150 – 30 – 15
Increment (L / Min)10.251
FloatBall – AISI 316Ball – PlasticBall – AISI 316
Inlet Connection3 / 8″ BSP RH / 1/8″ BSPT
Outlet Connection3 / 8″ BSP RH
FilterAISI 316 wire mesh
BodyBrass (Chrome plated)
Calibrated Pressure & Temperature Accuracy50 psig (3.45 bar) at210C / 60 psig (4.13 bar)at210C + 10% of the scale value or0.2 L/min whichever is higher
Humidifier Valve Setting 5 – 7 psig (34 – 48 kPa) auto resettable after bleeding

Note:
1. Read scale through the center of SS Ball (Float) and from the top of bobbin float for accuracy.
2. Thorpe Tube must be in upright position.
3. Medical Gas supply source is either through Outlet (DISS or Quick Connect Terminal Unit) of MGPS or Compressed Gas Cylinder
where pressure is suitably reduced by regulator.